6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Restart

Restart

The Restart function restarts the current sweep, or measurement, or set of averaged/held sweeps
or measurements. If you are Paused, pressing Restart does a Resume.

The Restart function is accessed in several ways:

Pressing the Restart key

Sending the remote command INIT:IMMediate

Sending the remote command INIT:RESTart

See

"More Information" on page 1000

Key Path

Front-panel key

Remote Command

:INITiate[:IMMediate]

:INITiate:RESTart

Example

:INIT:IMM

:INIT:REST

Notes

:INITiate:RESTart and :INITiate:IMMediate perform exactly the same function.

Couplings

Resets average/hold count k. For the first sweep overwrites all active (update=on) traces with
new current data. For application modes, it resets other parameters as required by the
measurement.

Status Bits/OPC
dependencies

This is an Overlapped command.

The STATus:OPERation register bits 0 through 8 are cleared.

The STATus:QUEStionable register bit 9 (INTegrity sum) is cleared.

The SWEEPING bit is set.

The MEASURING bit is set.

Backwards Compatibility
Notes

For Spectrum Analysis mode in ESA and PSA, the Restart hardkey and the INITiate:RESTart
command restart trace averages (displayed average count reset to 1) for a trace in Clear Write,
but did not restart Max Hold and Min Hold.

In the X-Series, the Restart hardkey and the INITiate:RESTart command restart not only Trace
Average
, but MaxHold and MinHold traces as well.

For wireless comms modes in ESA and PSA, the Restart hardkey and the INITiate:RESTart
command restart every measurement, which includes all traces and numeric results. There is no
change to this operation.

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference