Agilent Technologies Signal Analyzer N9030a User Manual

Page 554

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
BW

Example

BAND:VID:RAT 2

BAND:VID:RAT?

BAND:VID:RAT:AUTO 0

BAND:VID:RAT:AUTO?

Notes

The values shown in this table reflect the conditions after a Mode Preset.

Couplings

See  “Coupling Auto Rules”

Preset

1

ON

State Saved

Saved in instrument state

Min

0.00001

Max

3000000

Backwards Compatibility
Notes

For backwards compatibility this command obeys both the BANDwidth and BWIDth forms.

Initial S/W Revision

Prior to A.02.00

Modified at S/W
Revision

A.02.00

Auto Rules

The Auto Rules for the VBW:3dB RBW function follow.

First, if Source Mode is set to “Tracking”:  Use 1.0

Otherwise, we go through the following list of detector numbers and find the lowest numbered
detector being used on any active traces (traces for which Update is On):

1. Peak

2. Normal

3. Average

4. Sample

5. Negative Peak

6. EMI Average

7. Quasi Peak

8. RMS Average

Use that detector to pick the ratio based on the following criteria:

554

Remote Language Compatibility Measurement Application Reference

Advertising