Agilent Technologies Signal Analyzer N9030a User Manual

Page 1346

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference
Trace/Detector

Trace Display State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:DISPlay[:STATe] ON|OFF|0|1

:TRACe[1]|2|...6:DISPlay[:STATe]?

Example

TRAC2:DISP,1 Makes trace 2 visible

TRAC3:DISP,0 Blanks trace 3

Couplings

Whenever you set Update to On for any trace, the Display is set to On for that trace.

Preset

1|0|0|0|0|0 (On for Trace 1; Off for 2–6)

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRACe:MODE BLANk

sets :TRACe:UPDate OFF, :TRACe:DISPlay OFF, for the selected trace.  In earlier analyzers, View
and Blank were trace modes, set by TRACe:MODE command.  In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE BLANk command will yield its new equivalent, which is Update=Off, Display=Off 

Initial S/W Revision

Prior to A.02.00

More Information

When a trace becomes inactive, the following things happen:

•

Any update from the SENSe system (detectors) immediately stops (does not wait for end of

sweep)

•

the trace is displayed at half intensity (as long as it stays inactive)

Inactive traces display across the entire X Axis of the instrument. Their horizontal placement
does not change even if X Axis settings subsequently are changed, although Y-axis settings will
affect the vertical placement of data.

In most cases, inactive traces are static and unchanging; however, there are cases when an
inactive trace will update, specifically:

•

if data is written to that trace from remote

•

if trace data is loaded from mass storage

•

if the trace is the target of a Copy or participant in an Exchange

•

if the trace is cleared using the Clear Trace function (below)

When a trace becomes active (Update=On), the trace is cleared, the average count is reset, and a
new sweep is initiated.

1346