6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Trace/Detector

and Blank were trace modes, set by TRACe:MODE command.  In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE VIEW command will yield its new equivalent, which is Update=Off, Display=On

Initial S/W Revision

Prior to A.02.00

Trace Display State On/Off

Key Path

Trace/Detector

Remote Command

:TRACe[1]|2|...6:DISPlay[:STATe] ON|OFF|0|1

:TRACe[1]|2|...6:DISPlay[:STATe]?

Example

TRAC2:DISP,1 Makes trace 2 visible

TRAC3:DISP,0 Blanks trace 3

Couplings

Whenever you set Update to On for any trace, the Display is set to On for that trace.

Preset

1|0|0|0|0|0 (On for Trace 1; Off for 2–6)

State Saved

Saved in instrument state

Backwards
Compatibility SCPI

:TRACe:MODE BLANk

sets :TRACe:UPDate OFF, :TRACe:DISPlay OFF, for the selected trace.  In earlier analyzers, View
and Blank were trace modes, set by TRACe:MODE command.  In the X-Series, View and Blank
are two of the states set by the :TRACe:UPDate and :TRACe:DISPlay commands. The
TRACe:MODE BLANk command will yield its new equivalent, which is Update=Off, Display=Off 

Initial S/W Revision

Prior to A.02.00

More Information

When a trace becomes inactive, the following things happen:

Any update from the SENSe system (detectors) immediately stops (does not wait for end of

sweep)

the trace is displayed at half intensity (as long as it stays inactive)

Inactive traces display across the entire X Axis of the instrument. Their horizontal placement
does not change even if X Axis settings subsequently are changed, although Y-axis settings will
affect the vertical placement of data.

In most cases, inactive traces are static and unchanging; however, there are cases when an
inactive trace will update, specifically:

if data is written to that trace from remote

if trace data is loaded from mass storage

Remote Language Compatibility Measurement Application Reference