6  RLC Swept SA Measurement Front-Panel & SCPI Reference
View/Display

Key Path

View/Display, Display, System Display Settings, Backlight

Example

DISP:BACK ON

Readback

On

Initial S/W Revision

Prior to A.02.00

Off

Turns the display backlight off.

Key Path

View/Display, Display, System Display Settings, Backlight

Example

DISP:BACK OFF

Readback

Off

Initial S/W Revision

Prior to A.02.00

Backlight Intensity

An active function used to set the backlight intensity. It goes from 0 to 100 where 100 is full on
and 0 is off. This value is independent of the values set under the Backlight on/off key.

Key Path

View/Display, Display, System Display Settings

Remote Command

:DISPlay:BACKlight:INTensity <integer>

:DISPlay:BACKlight:INTensity?

Example

DISP:BACK:INT 50

Preset

100 (Set by Restore Misc Defaults)

Min

0

Max

100

Initial S/W Revision

Prior to A.02.00

Normal

Single window view of the frequency domain or zero span. This is the classic SA view. This is
also the View into which the analyzer switches whenever you do anything that causes the
frequency limits to change, for example:

If you switch inputs (for example, if you switch from the RF Input to External Mixing)

If, while in External Mixing, you edit the Harmonic Table

1454

Remote Language Compatibility Measurement Application Reference