5  Legacy Command Descriptions
CA (Couple Attenuation)

CA (Couple Attenuation)

Syntax

Legacy Products

8560 series, 8566A/B, 8568A/B

Description

During normal operation, the instrument’s input attenuation is coupled to the reference level. This
coupling keeps the mixer input at a level such that a continuous wave signal displayed at the
reference level is at or below –10 dBm (or the value specified in the ML command.)

The CA command sets the threshold to –10 dBm (or to the value specified by the commands

"ML

(Mixer Level)" on page 411

or

"KS, (Mixer Level)" on page 318

). The counterpart to the CA

command is

"AT (Input Attenuation)" on page 236

, which allows levels less than the threshold

value at the mixer input.

Format

CA

Query Data Type

N/A

SCPI Equivalent
Commands

[:SENse]:POWer[:RF]:ATTenuation:AUTO ON (see

"(Mech) Atten " on page 519

)

254

Remote Language Compatibility Measurement Application Reference