6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Meas Setup

Demod State  (Remote Command Only)

Sets or queries the state of the Analog Demod Tune and Listen function. Setting the state to ON
with this command will select AM demodulation by default and activate it (turn it on).

The response to the query is determined by the current setting of [:SENSE]:DEMod
AM|FM|PM|OFF. The response will be 1 if AM, FM, PM are selected, or 0 if OFF is selected..

Remote Command

[:SENSe]:DEMod:STATe OFF|ON|0|1

[:SENSe]:DEMod:STATe?

Preset

OFF

Initial S/W Revision

Prior to A.02.00

Noise Source

This menu allows you to turn the noise source power on or off when making manual noise figure
measurements. 

See

"More Information" on page 854

.

Key Path

Meas Setup

Remote Command

:SOURce:NOISe:TYPE NORMal | SNS

:SOURce:NOISe:TYPE?

Example

SOUR:NOIS:TYPE NORM

Couplings

If no SNS is connected, this parameter will be set to “Normal”

When Type is set to “SNS” and the SNS is disconnected, this parameter gets bumped to
“Normal”

When an SNS is not connected, the SNS type will be grayed (disabled).

Preset

Normal

State Saved

Saved in instrument state.

Range

Normal | SNS

Backwards Compatibility
Notes

In previous Noise Figure analysis applications, this command

could optionally be preceded with the :SENSe keyword.  The

optional :SENSe keyword is no longer supported.

Initial S/W Revision

Prior to A.02.00

Remote Language Compatibility Measurement Application Reference