Cmd error – Agilent Technologies Signal Analyzer N9030a User Manual

Page 887

Advertising
background image

6  RLC Swept SA Measurement Front-Panel & SCPI Reference

Mode Setup

Selecting any legacy analyzer from this menu performs an instrument preset and sets Span, Trace
Points, couplings, VBW/RBW ratio, and Span/RBW ratio appropriately, as shown in the table
below.

Remote
Language

Start
Freq.

Stop
Freq.

Number of Trace
Points

RF
Coupling

VBW/RBW Ratio

Span/RBW
Ratio

8560E/EC

30 Hz

2.9 GHz

601

AC

1

91

8561E/EC

30 Hz

6.5 GHz

601

AC

1

91

8562E/EC

30 Hz

13.2
GHz

601

AC

1

91

8563E/EC

30 Hz

26.5
GHz

601

DC

1

91

8564E/EC

30 Hz

40 GHz

601

DC

1

91

8565E/EC

30 Hz

50 GHz

601

DC

1

91

HP8566A

2 GHz

22 GHz

1001

DC

3

(VBW one step wider
than RBW)

106

HP8566B

2 GHz

22 GHz

1001

DC

3

(VBW one step wider
than RBW)

106

HP8568A

0 Hz

1.5 GHz

1001

DC

3

(VBW one step wider
than RBW)

106

HP8568B

0 Hz

1.5 GHz

1001

DC

3

(VBW one step wider
than RBW)

106

Span, Trace Points, Couplings, VBW/RBW Ratio, and Span/RBW Ratio Settings

Key Path

Mode Setup

Cmd Error

Turning Cmd Error On or Off enables or disables the display of the "CMD ERR" error messages. The
default setting is On. The error message appears in the Message bar and also can be queried using
ERR?

. The error message occurs if either the command syntax or any of its parameters are

incorrectly formed. The selected value is preserved after presetting or power cycling the
instrument. Disabling the display of command errors disables the display of all error types.

The format of the errors are as follows:

1. CMD ERR, <string>

Remote Language Compatibility Measurement Application Reference

887