5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1012

■■■■■■■■

tiamo 2.3 Patch 1

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously.

Range

-20.0 ... 150 °C

Default value

25.0

5.6.4.9.4

MEAS Conc - Additional measured values

Tab: Method

Commands MEAS Conc Properties... Additional measured

values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together with the measured values present in
the default settings in additional measured value columns Calculated #
or External # in the measuring point list.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.

[New]

Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 930)
.

[Properties]

Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be edited (see Chapter
5.6.3.11.1, page 930)
.

[Delete]

Delete the calculated measured value selected in the table.