5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

740

■■■■■■■■

tiamo 2.3 Patch 1

Designation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement fulfil-
led
, 0 = Condition requirement not fulfilled

.CONC

Concentration of the solution used for the command

.CYL

Cylinder volume of the exchange or dosing unit used
for the command

.DBL

Total duration for the processing of the command in s

.DRI

Current and/or last drift for drift correction in µL/min

.DSC

Time for processing all start conditions s

.DTI

Time for drift correction (time from the start of the
titration to the end of the command) in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.EP{x}.MEA

Measured value for the endpoint × (1…9) in the unit of
the measured value

.EP{x}.MEP

Quantity of endpoints in the window × (1…9); 1 = 1
endpoint
, 2 = 2 or more endpoints, 3 = EP cor-
rected with Autodrift, 4 = EP corrected with
manual drift

.EP{x}.TEM

Temperature for the endpoint × (1…9) in °C

.EP{x}.TIM

Time in s until the endpoint × (1…9) is reached

.EP{x}.VOL

Volume for the endpoint × (1…9) in mL

.ETE

End temperature (temperature following processing of
the command) in °C

.EVT

End volume (total dosed volume at the end of the
command) in mL

.FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) = Com-
mand has never ended

.FP{x}.MEA

Measured value for the fixed endpoint × (1…9) in mV