Metrohm tiamo 2.3 Patch 1 User Manual

Page 60

Advertising
background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

44

■■■■■■■■

tiamo 2.3 Patch 1

Designation

Description

Commands

.ETR

Result of the electrode test.

Electrode quality:

0 = No data available

1 = Short circuit

2 = Excessive start drift

3 = Wrong buffer

4 = Problems with the diaphragm

5 = Reference system defective

6 = No buffer values for the tem-
perature or excessive deviation of
the temperature between buffers

7 = Partial short circuit

8 = Glass defective

9 = Poor electrode

10 = Usable electrode

11 = Good electrode

12 = Excellent electrode

ELT LOOP

.EVT

End volume (total dosed volume
at the end of the command) in mL

DET, MET, SET,
KFT, STAT, DOS

.FIN

Command status; 1 = Command
has ended at least once
, 0,
invalid (variable not available) =
Command has never ended

All except START
and END

.FP{x}.CNT

Intensity for the fixed endpoint x
(1…9) in counts

MEAS Opt

.FP{x}.DME

Measured value differential for the
fixed endpoint x (1…9)

MET

.FP{x}.DVT

Drift for the fixed endpoint x (1…
9) in µg/min

KFC, BRC

.FP{x}.ERC

ERC for the fixed endpoint x (1…
9)

DET

.FP{x}.MEA

Measured value for the fixed end-
point x (1…9) in mV

DET, MET, SET,
KFT, STAT