5 Method

tiamo 2.3 Patch 1

■■■■■■■■

827

Titration parameters
Parameters for the run of the titration.

Stop conditions
Entry of the stop time which causes the titration to stop.

Conditioning
Enable/Disable the conditioning. Definition of start drift, drift correction
and drift value.

Additional evaluations
Definition of additional methods for the evaluation of the titration
curves.

Additional measured values
Definition of additional measured values of other measuring com-
mands which are saved as additional columns in the measuring point
list.

Command variables
The following command variables are generated in the method run by the
command KFC and can be used in formulas under the designation 'Com-
mand name.Variable designation'
:

Designation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.CBY

Command status; 1 = Conditioning activated, 0 =
Conditioning deactivated

.COK

Command status; 1 = Condition requirement fulfil-
led
, 0 = Condition requirement not fulfilled

.DBL

Total duration for the processing of the command in s

.DRI

Current and/or last drift for drift correction in µL/min

.DTI

Time for drift correction (time from the start of the
titration to the end of the command) in s

.EME

End measured value (measured value after processing
of the command) in the unit of the measured value

.EP.CHA

Charge for endpoint in mA·s

.EP.DVT

Drift for the endpoint in µg/min

.EP.MEA

Measured value for the endpoint in mV

.EP.QTY

Measured value (water) for the endpoint in µg

.EP.TEM

Temperature for the endpoint 1 in °C