Metrohm tiamo 2.3 Patch 1 User Manual

Page 58

Advertising
background image

2.4 Formula editor

■■■■■■■■■■■■■■■■■■■■■■

42

■■■■■■■■

tiamo 2.3 Patch 1

Designation

Description

Commands

.DBL

Total duration for the processing
of the command in s

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS, CAL
Cond, CAL MEAS,
CAL Spec,
STDADD, DOS,
ELT MEAS

.DEF

Default value which is defined in
the configuration for an IO port

ANALOG OUT,
DIGITAL OUT

.DRI

Current and/or last drift for drift
correction in µL/min

SET, KFT, KFC,
BRC

.DSC

Time for processing all start condi-
tions s

DET, MET, SET,
KFT, STAT

.DTI

Time for drift correction (time
from the start of the titration to
the end of the command) in s

SET, KFT, KFC,
BRC

.EGF

Last measured gas flow rate
(measured value following pro-
cessing of the command) in
mL/min

MEAS T/Flow

.EME

End measured value (measured
value after processing of the com-
mand) in the unit of the measured
value

DET, MET, SET,
KFT, KFC, BRC,
STAT, MEAS
(without Ref,
Spec), CAL Cond,
CAL MEAS,
STDADD, DOS,
ELT MEAS

.ENP

Electrode zero point for the sensor
used for the command (dimen-
sionless for pH sensor or in mV for
the ISE sensor) and/or electrode
zero point calculated from the cal-
ibration (for STDADD and CAL
LOOP)

DET pH, DET U,
MET pH, MET U,
SET pH, SET U,
STAT, MEAS pH,
MEAS U, MEAS T,
MEAS Conc,
STDADD, CAL
LOOP, DOS, ELT
LOOP

.EP{x}.CHA

Charge for endpoint x (1…9) in
mA·s

KFC, BRC