4 cal cond - conductivity standard – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1117

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1101

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Temperature

Measurement temperature

Temperature of the solution to be measured. If a temperature sensor is
connected, the temperature entered here is ignored.

Range

-20.0 ... 150.0 °C

Default value

20.0

5.6.5.4.4

CAL Cond - Conductivity standard

Tab: Method

CAL Cond Properties... Conductivity standard

Command name

Name of the command.

Input

25 characters

Definition of the conductivity standard
Selection of the type as to how the data are to be defined for the conduc-
tivity standard.

Selection

Input | Request | Template

Default value

Input

Input
If this option is selected, then the values for the conductivity standard
can be entered directly.
Request
If this option is selected, then first the current temperature of the solu-
tion is measured after the start of the determination. Afterwards, the
dialog window Data Conductivity standard will appear (see Chap-
ter 5.6.5.4.5, page 1102)
. The method continues running after the
data has been entered.
Template
If this option is selected, then a conductivity standard can be selected,
and its data will be used to calculate the temperature coefficient for
the specified reference temperature.

Parameters for input
If the option Input has been selected, then the following parameters will
be displayed: