2 cal cond - general/hardware – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1112

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1096

■■■■■■■■

tiamo 2.3 Patch 1

Identifica-
tion

Description

.DBL

Total duration for the processing of the command in s

.EME

Final measured value (measured value after processing
of the command) in the unit of the measured value

.ETE

End temperature (temperature following processing of
the command) in °C for the last measurement

.FIN

Command status: 1 = Command has been ended at
least once
, 0 = Command was started but not
ended
, invalid = Command was not started (varia-
ble not available)
.

.IME

Initial measured value (measured value at the start of
the command) in the unit of the measured value for the
last measurement

.ITE

Initial temperature (temperature before processing of
the command) in °C for the last measurement

.LP.MEA

Measured value for the last measuring point in the
measuring point list in the unit of the measured value

.LP.TEM

Temperature for the last measuring point of the measur-
ing point list in °C

.LP.TIM

Time in s until the last measuring point in the measuring
point list is reached

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in the measuring point list

.RTE

Reference temperature in °C (only MEAS Cond and
CAL Cond)

.STY

Type of stop with which the command was stopped: 1
= stopped normally
, 0 = stopped manually with
[Quit]
or [Stop].

.TC

Temperature coefficient in %/°C

5.6.5.4.2

CAL Cond - General/Hardware

Tab: Method

CAL Cond Properties... General/Hardware

Command name

Name of the command.

Input

25 characters

Advertising