Metrohm tiamo 2.3 Patch 1 User Manual

Page 1141

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1125

Moves concentration upward (modifies sequence).

Moves concentration downward (modifies sequence).

[New]

Add the concentration for the next standard.

[Properties]

Edit the concentration of the selected standard.

[Delete]

Delete the selected concentration.

Request for standard exchange

on | off (Default value: on)

If this check box is activated, a request for manual solution exchange is
displayed after each time the calibration loop is run through. This option
must be disabled for automatic solution exchange when calibrating with a
Sample Processor.

Calibration curve

Calibration function

Selection of the function with which the calibration curve is to be calcula-
ted out of the calibration points.

Selection

Linear | Quadratic | Cubic | automatic

Default value

Linear

Confidence interval

Range within which the measured values for the determination of the cali-
bration curve must lie. Values outside this range are called outliers, they
are not integrated in the calculation of the calibration coefficient.

Selection

99 % | 98 % | 95 % | off

Default value

98 %

off
The confidence interval will not be checked.

Advertising