6 meas ipol - fixed endpoint evaluation – Metrohm tiamo 2.3 Patch 1 User Manual

Page 989

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

973

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 931)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 931)
.

[Delete]

Delete the external measured value selected in the table.

5.6.4.5.6

MEAS Ipol - Fixed endpoint evaluation #

Dialog window: Method

MEAS Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation #

Measured quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time

Default value

Measured value

Fixed value

Value of the fixed endpoint.

Measured value (Titrando, 867, 855)
Range

-2000.0 ... 2000.0 mV

Measured value (Titrino)
Range

-2000 ... 2000 mV

Time
Range

0.0 ... 999999.9 s