11 det ipol - fixed endpoint evaluation, 5 det upol, 1 det upol - overview – Metrohm tiamo 2.3 Patch 1 User Manual

Page 555: Det upol

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

539

5.6.3.2.4.11

DET Ipol - Fixed endpoint evaluation

Dialog window: Method

DET Ipol Properties... Additional evalua-

tions

Fixed endpoint evaluation #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-2000.0 ... 2000.0 mV

Time
Range

0.0 ... 999999.9 s

Volume
Range

0.00000 ... 9999.99 mL

5.6.3.2.5

DET Upol

5.6.3.2.5.1

DET Upol - Overview

Dialog window: Method

DET Upol Properties... DET Upol - 'Command

name'

Command for Dynamic equivalence point titrations with amperomet-
ric measurement (selectable polarization voltage).

Instruments
This command can be executed with the following devices.

Titrando: 808, 809, 835, 836, 857, 888, 904, 905, 906, 907

Titrino: 716, 721, 736, 751, 785, 794, 798, 799

Robotic Titrosampler: 855

Appearance
The command has the following appearance: