6 stat ph - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 User Manual

Page 892

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

876

■■■■■■■■

tiamo 2.3 Patch 1

Start time

The measuring points are not entered in the measuring point list until this
time has elapsed since the start of the titration.

Range

0 ... 999999 s

Default value

0 s

Start meas. value pH

The measured values are not entered in the measuring point list until this
value has been reached.

Titrando, 855
Range

-20.000 ... 20.000

Selection

off

Default value

off

Titrino
Range

-20.00 ... 20.00

Selection

off

Default value

off

Start rate

The measured values are not entered in the measuring point list until the
dosing rate has fallen below this value. This parameter is not checked until
10 s after the start.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

off

Default value

off

Titrino
Range

0.01 ... 150.00 mL/min

Selection

off

Default value

off

5.6.3.8.6.6

STAT pH - Stop conditions

Tab: Method

STAT pH Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Advertising