5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

970

■■■■■■■■

tiamo 2.3 Patch 1

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.5.6, page 973).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the minimum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

1.0 ... 2000.0 mV/s

Default value

25.0 mV/s

Maximum evaluation

Maximum evaluation

on | off (Default value: off)

If this option is enabled, then the associated volume, time and tempera-
ture for the maximum measured value will be interpolated from the meas-
uring point list.

Threshold value

The evaluation of the maximum begins as soon as the slope of the curve
exceeds the set threshold value.

Range

1.0 ... 2000.0 mV/s

Default value

25.0 mV/s

Break point evaluation

Break point evaluation

on | off (Default value: off)

If this option is activated, then sharp, almost 90° changes in directions in
the measurement curve will be evaluated.