4.5 Determination overview

■■■■■■■■■■■■■■■■■■■■■■

346

■■■■■■■■

tiamo 2.3 Patch 1

ted from these measured values. Listed below are the individual measured
values and the results for Slope, E(0), c(blank) and Variance.

Calibration curve for CAL LOOP Opt
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration curve and calibration
data will be displayed for each CAL LOOP Opt command. The command
type is shown above the calibration curve. The curve display shows the
measured values and the curve calculated from these measured values.
Listed below are the individual measured values and the results: conc
['Unit']
,

ε[mAU], T [°C], t [s], Standard #, Wavelength, c0, c1, c2,

c3, Confidence interval, Number of outliers and Calibration range.

Standard addition curve for STDADD
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the standard addition curve and evalu-
ation data will be displayed for each STDADD command (see Chapter
5.6.4.17.1, page 1066)
. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calcula-
ted from these measured values. Listed below are the individual measured
values and the results for Determination, Show calibration curve...,
E(0), Determination, Show calibration curve... and Variance.

Calibration data for CAL Cond
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration data will be displayed
for each CAL Cond command (see Chapter 5.6.5.4.1, page 1095). Dis-
played are the command type, the measured values and the cell constant
calculated with these measured values.

Calibration curve for ELT LOOP
In the dialog window Calibration curve/Calibration data a tab marked
with the command name showing the calibration curve and calibration
data will be displayed for each CAL Cond command (see Chapter
5.6.5.8, page 1115)
. The command type is shown above the calibration
curve. The curve display shows the measured values and the curve calcula-
ted from these measured values. Listed below are the individual measured
values and the results for Slope, pH(0) and Variance.

Note

Calibration curves cannot be printed out directly from the dialog win-
dow Calibration curve/Calibration data. Instead you should use a
report with a report template which contains the fixed report (see
Chapter 4.4.1.4.3.13, page 250)
.