5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1185

exchange unit or dosing device used. If the entered filling rate is too high
for the selected dosing device, it will automatically be reduced during dos-
ing to the largest possible value.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Sensor

Measuring input

Selection of the measuring input to which the sensor is connected.

Titrando
Selection

1 | 2

Default value

1

855
Selection

1

Default value

1

Titrino
Selection

1 | 2 | diff.

Default value

1

Sensor

Selection of a sensor of type pH electrode, metal electrode, pH elec-
trode
or ISE electrode from the list of sensors defined in the sensor
table. The calibration data for the sensor will be adopted for pH and ISE
electrodes.

Selection

Sensor name | pH electrode | Metal electrode |
ISE electrode

Default value

Metal electrode

Temperature measurement

Type of temperature measurement.

Titrando, 855
Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.