Metrohm tiamo 2.3 Patch 1 User Manual

Page 642

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

626

■■■■■■■■

tiamo 2.3 Patch 1

command, a check is made of the working life, the validity of the titer and
the GLP test interval for the selected solution.

Input

24 characters

Selection

'Solution name' | not defined

Default value

not defined

not defined
No tests will be carried out.

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando (without 888)
Selection

1 | 2

Default value

1

855, 888
Selection

1

Default value

1

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

I(pol)

The polarization voltage is the voltage applied to the polarizable electrode
during a amperometric measurement.

Titrando, 855
Range

–125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

–127 ... 127 µA (Increment: 1)

Default value

5 µA

Electrode check

on | off (Default value: off)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status