5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

836

■■■■■■■■

tiamo 2.3 Patch 1

756, 831
Range

0.3 ... 999.9 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

851, 852
Range

0.3 ... 999.0 µg/min

Default value

0.3 (slow), 15.0 (optimal), 30.0 (fast) µg/min

Selection

minimum

Stop criterion

Stop criterion

Switches off titration when the endpoint has been reached and the stop
criterion has been met.

756, 831
Selection

Drift | rel. drift

Default value

rel. drift

851, 852
Selection

Drift | rel. drift | Time | off

Default value

rel. drift

Stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.

Range

1 ... 999 µg/min

Default value

5 µg/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Range

0 ... 999 µg/min

Default value

5 µg/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time.

Range

0 ... 999 s

Default value

10 s