8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1606

■■■■■■■■

tiamo 2.3 Patch 1

Titrando, 855
Range

–125.0 ... 125.0 µA (Increment: 0.5)

Default value

5.0 µA

Titrino
Range

–127 ... 127 µA (Increment: 1)

Default value

5 µA

U(pol)

The polarization voltage is the potential applied to the polarizable elec-
trode during an amperometric measurement.

Titrando, 855
Range

–1250 ... 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Range

–1270 ... 1270 mV (Increment: 10)

Default value

400 mV

Reference temperature

The electrical conductivity depends greatly on the temperature. The con-
ductivity measured at a particular temperature is thus automatically con-
verted to this reference temperature (usually 20 °C or 25 °C).

712
Range

–170.0 ... 500.0 °C

Default value

20.0

856
Range

–20.0 ... 150.0 °C

Default value

20.0

Selection

off

off
The conductivity at the measuring temperature is displayed.

Temperature compensation
Selection

Temperature coefficient | Sample solution

Default value

Temperature coefficient

Temperature coefficient
Manual entry of a constant temperature coefficient.
Sample solution
Selection of a function for the temperature coefficient from the table
Sample solution (TC conductivity).