5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

942

■■■■■■■■

tiamo 2.3 Patch 1

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s

Stop measured value pH

Stops when the preset measured value has been reached since the start of
the measurement.

Titrando, 855, 867
Range

-20.000 ... 20.000

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Titrando, 855, 867
Range

0 ... 999999 s

Default value

120 s

Titrino
Range

0 ... 9999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

Titrando, 855, 867
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s (Increment: 0.08)

Default value

2.0 s