5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1275

Note

If the run should not be stopped (for example for back-weighing in
KF titrations), the REQUEST command must be inserted in a track
other than the one with the titration command with the option Return
immediately
.

Sample position

on | off (Default value: off)

If this option is enabled, then the position of the sample on the rack will
be queried.

ID1 … ID16

on | off (Default value: off)

If this option is enabled, then the sample identifications ID1 … ID16 will
be requested. If other names were assigned to this variables in the START
command, these these names will be displayed in the dialog window
Sample data request.

Note

Text is selected as type in the default settings for the sample identifica-
tions ID1 … ID16. If you wish to enter numbers for these method vari-
ables for use in later calculations, then these variables must be switched
in the dialog window Method variable - Sample position to the
type number. START

Properties... Method variables [Prop-

erties]

Method variables - Sample position

Sample size

on | off (Default value: off)

If this option is switched on, then the sample size will be queried.

Sample size unit

on | off (Default value: off)

If this option is switched on, then the sample size unit will be queried.

Message
on | off (Default value: off)

If this option is enabled, then the message defined here will appear in the
dialog window Sample data request in which the sample data is quer-