5 Method

tiamo 2.3 Patch 1

■■■■■■■■

473

Max. increment

A maximum volume increment should be selected when the titrant con-
sumption up to the equivalence point is expected to be very small, a start
volume is to be added until just before the equivalence point is reached or
if the change of direction in the potential jump region is very abrupt, as
otherwise it is easy to add too large a volume in the equivalence point
region. The value should not be less than 1/100 cylinder volume.

Titrando, 855
Range

0.1 ... 9999.9 µL

Selection

off

Default value

off

Dosing rate

Speed at which the volume increments are added. The maximum dosing
rate depends on the cylinder volume of the exchange unit or dosing unit
used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Temperature

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C