5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1211

Move

Selecting the target position on the rack:

Selection

Sample position | Rack position | Special
beaker
| Relative angle

Default value

Sample position

Sample position
Position defined for the current determination in the Run subwindow
of the program part Workplace as the Sample position parameter.
Rack position
Absolute position on the rack.
Special beaker
Special position on the rack defined in the rack properties of the Con-
figuration
program part.
Relative angle
The sample rack can be rotated by a certain angle relative to the cur-
rent position independently of rack positions. This can be used for
example to remove vial caps automatically.

Number

Specification of the number of the rack position. This parameter is only
visible for Move = Rack position or Special beaker. The rack position
can be entered as a formula as well, which is necessary for the automatic
calibration of electrodes, for example. The formula editor is opened by
right-clicking on the input field.

Move = Rack position
Range

1 ... 999

Default value

1

Move = Special beaker (for 778, 789, 814, 815, 855, 864, 874)
Range

1 ... 16

Default value

1

Move = Special beaker (for 730)
Range

1 ... 8

Default value

1

Move = Special beaker (for 774)
Input

1

Move angle

This parameter specifies the angle by which the sample rack should be
rotated relative to the current position. It is only visible for Move = Rela-
tive angle
.