5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1243

Display message

on | off (Default value: on)

If this option is enabled, all active tracks are stopped and the message
defined in the text box will be displayed in the event that the limits are
exceeded. You can select whether you want to continue the stopped
tracks with [Continue] or - if defined - to execute the Action listed
below.

Record message

on | off (Default value: on)

If this option is enabled, the message defined in the text box will be docu-
mented in the determination if the limits are exceeded. The run will not be
stopped.

Message by E-mail

on | off (Default value: off)

If this option is enabled, the message defined in the text box will be sent
to the address defined under [E-mail] in the event that the result limits
are exceeded.

[E-mail]

Opens the dialog window Send E-mail for defining the E-mail parameters
(see Chapter 2.6.1, page 95).

Acoustic signal

on | off (Default value: off)

If this option is enabled, an acoustic signal will be emitted in addition to
the message defined above when a limit is exceeded.

Action

Action

on | off (Default value: off)

If this option is enabled, one of the following actions will be executed in
the event that limit values are exceeded.

Selection

Cancel determination | Stop determination
and series

Default value

Cancel determination

Cancel determination
The determination is stopped and the next determination of the series
is started.
Stop determination and series
Both the determination and the series are stopped.