Metrohm tiamo 2.3 Patch 1 User Manual

Page 969

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

953

Measurement
The two measuring modes Measurement with drift control or Meas-
urement without drift control
can be selected.

Selection

Measurement with drift control | Measure-
ment without drift control

Default value

Measurement with drift control

Measurement with drift control
The measurement is carried out drift-controlled if this option is ena-
bled. The measurement is stopped as soon as the defined Signal drift
or the Stop measured value is reached or a waiting time has
elapsed. The corresponding parameters are not visible if this option is
disabled.
Measurement without drift control
The measurement is carried out without drift control if this option is
enabled. The measurement continues until one of the stop criterions
Measuring time or Stop measured value is met. The correspond-
ing parameters are not visible if this option is disabled.

Note

A constant measured value is often only reached after a certain time, as
mixing and possibly the reaction itself require a certain time. The
response time of an electrode can also increase with time, i.e. reaching
a constant measured value takes longer and longer. Drift-controlled
measurement
is particularly advisable in such cases, as the measured
values are not applied until equilibrium has almost been reached.

Measurement with drift control
If the option Measurement with drift control has been selected, then
the following parameters will be displayed:

Measured value drift

The measured value is only accepted if the signal drift defined here has
been fallen short of.

Titrando, 855, 867
Range

0.1 ... 999.0 mV/min

Default value

10.0 mV/min

Titrino
Range

0.5 ... 999.0 mV/min

Default value

10.0 mV/min