5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1166

■■■■■■■■

tiamo 2.3 Patch 1

Identifica-
tion

Description

.FIN

Command status; 1 = Command has been ended
at least once
, 0, invalid (variable not available) =
Command has never been ended

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the measured
value

.ITE

Initial temperature (temperature before start conditions
are processed) in °C

.LP.CAx

Calculated value × (1...3) for the last measuring point
of the measuring point list

.LP.DVT

dV/dt for the last measuring point in the measuring
point list (SET, KFT, STAT, DOS) or drift for the last
measuring point in in the measuring point list in
µg/min (KFC)

.LP.EXx

External value × (1...3) for the last measuring point of
the measuring point list

.LP.MEA

Measured value for the last measuring point in the unit
of the measured value of the measuring point list

.LP.TEM

Temperature for the last measuring point of the meas-
uring point list in °C

.LP.TIM

Time in s until the last measuring point in the measur-
ing point list is reached

.LP.VOL

Volume for the last measuring point in the measuring
point list in mL

.MR.MRC

Correlation coefficient for mean dosing rate for the
whole range

.MR.MRS

Standard deviation for mean dosing rate for the whole
range in mL/min

.MR.MRT

Mean dosing rate for the whole range in mL/min.

.MTE

Temperature measurement with sensor; 1 = on, 0 =
off
)

.NMP

Number of measuring points in the measuring point list

.SLO

Electrode slope of the sensor used for the command (in
%)