6 Configuration

tiamo 2.3 Patch 1

■■■■■■■■

1381

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Maximum dosing/filling rate for the exchange unit, depending on
the cylinder volume:

Cylinder volume

maximum rate

1 mL

3.0 mL/min

5 mL

15.0 mL/min

10 mL

30.0 mL/min

20 mL

60.0 mL/min

50 mL

150.0 mL/min

Note

Enter lower rates for high-viscosity liquids.

Tubing parameters
Definition of the length and diameter of the tubing connected to the
exchange unit.

Note

Default values have already been entered for the tubing parameters;
these correspond to the dimensions of the standard tubing supplied. If
you do not make any modifications to the tubing connections then you
do not need to modify the tubing parameters. For information on
length and diameter of other tubing see Products/Accessories and
columns
under

http://products.metrohm.com

.

Dosing tip

Length

Length of the tubing connecting the cock to the dosing tip.

Range

0.0 ... 999.9 cm

Default value

40.0 cm

Diameter

Diameter of the tubing connecting the cock to the dosing tip.