Range
0.01 ... 150.00 mL/min
Selection
maximum
Default value
maximum
Maximum dosing/filling rate for the exchange unit, depending on
the cylinder volume:
Cylinder volume
maximum rate
1 mL
3.0 mL/min
5 mL
15.0 mL/min
10 mL
30.0 mL/min
20 mL
60.0 mL/min
50 mL
150.0 mL/min