Metrohm tiamo 2.3 Patch 1 User Manual

Page 30

Advertising
background image

1.10 What is new in tiamo 2.3?

■■■■■■■■■■■■■■■■■■■■■■

14

■■■■■■■■

tiamo 2.3 Patch 1

In the command START, Number or Date/Time was selected on the
tab Method variables for IDx as the type, while at the same time the
check box Check at start was also active. If the IDx fields were still
empty at the start of the determination, then the message 014-152
Field empty
would appear after the start of the conditioning.

Determinations with the command STAT pH with monitoring (action:
Wait for [Continue]) did not function correctly.

The Round() function supplied incorrect results under certain condi-
tions.

Formula

Result

Round( 1000 * 10 / 50 )/1000

0.00

wrong

(Round( 1000 * 10 / 50 )/1000)

0.20

right

If the iConnect was disconnected during a measurement, then the
message Unhandled error would appear.

If a method was comprised of several tracks with WAIT commands in
which one track called up the next with a command CALL, then this
sometimes led to deadlocks.

Program part Database

The table was not displayed in the dialog window Reprocessing on
the tab Statistics (the error was rectified previously in tiamo 2.2 Patch
1).

Problems sometimes arose in connection with titrations with condition-
ing in cases where more than one database was entered in the com-
mand DATABASE.

If a track (e.g. which contained a command STAT pH) was run
through more than once, then all of the curves were displayed in the
database, even if the function Delete old data was activated.

The break points found were not output in the database in the sub-
window Results on the tab Raw data with the command MEAS
Opt
.

The x smoothing was not applied on the y1 axis in the database with
the properties of the curve in the dialog window Curve evaluation.

If the sample size was changed and a determination was recalculated,
then even though the modification was applied everywhere in the
determination, this was still not the case in the statistical overview of
the report.

Method variables were always checked during recalculation, even if the
setting Check at start was not activated (command START, tab
Method variables).

Crashes sometimes occurred during recalculation due to multithreading
problems.