5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1174

■■■■■■■■

tiamo 2.3 Patch 1

Selection

off

off
Means that no stop will take place.

Stop time

Stops when the preset time has elapsed since the start of the dosing.

Range

0 ... 999999 s

Selection

off

Default value

off

off
Means that no stop will take place.

Filling rate

Rate at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used. If the entered filling rate is too high for the selected
dosing device, it will automatically be reduced during dosing to the largest
possible value.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Note

The filling rate should be reduced for viscous liquids.

5.6.6.3.4.5

DOS pH - Monitoring

Tab: Method

DOS pH Properties... Monitoring

Command name

Name of the command.

Input

25 characters

The following monitoring processes can be switched on and defined on
this tab: