5 Method

tiamo 2.3 Patch 1

■■■■■■■■

471

fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration
rate.

Measured value acceptance

Note

Will be displayed only if the option User is selected in the selection list
Titration rate.

Measured value drift

Drift for the measured value acceptance during the titration.

Titrando, 855
Range

0.1 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Titrino
Range

0.5 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Titrando, 855
Range

0 (slow, optimal, fast) ... 999999 s

Default value

0 (slow, optimal, fast) s