4 meas opt conc - additional measured values – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1075

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1059

Range

0.25 ... 999999.0 s

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off

Temperature (089 Photometer)

Temperature

Measuring temperature which can be entered manually. If the Tempera-
ture measurement
is set on the General/Hardware tab under Sensor
to continuous, then the temperature will be measured continuously.

Range

20.0 ... 60 °C

Default value

25.0 °C

5.6.4.15.4

MEAS Opt Conc - Additional measured values

Tab: Method

MEAS Opt Conc Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings.

Additional calculated measured values

Additional calculated measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be calculated from existing measured values and saved, which can also be
presented as curves under the designation Calculated 1…3 and used in
formulas as variables 'Command name.CA1…3'.

[New]

Open the dialog window Calc. measured value # in which the parame-
ters for the additional measured value can be entered (see Chapter
5.6.3.11.1, page 930)
.