8 Manual control

tiamo 2.3 Patch 1

■■■■■■■■

1587

8.4.1.5

Dosing - Emptying

Program part: Manual control

Here the emptying of dosing units can be started and stopped. This tab
appears when selecting a single dosing device as well as with all dosing
devices with dosing units.

Note

This tab is not displayed for external dosing devices on the Titrinos
736, 751, 758, 799
.

[Start]

Start emptying for the selected dosing device(s). The parameters defined
for the dosing units will be used.

[Stop]

Stop emptying for the selected dosing device(s).

8.4.1.6

Dosing - Dosing a fixed volume

Program part: Manual control

Here the dosing of a preset volume can be started and stopped. This tab
only appears if a single dosing device is selected.

Volume

Fixed volume that is to be dosed.

Range

0.100 ... 99999.9 mL

Default value

0.100 mL

Dosing rate

The volume is dosed at this rate. The maximum dosing rate depends on
the cylinder volume of the exchange unit or dosing unit used. If the
entered dosing rate is too high for the selected dosing device, it will auto-
matically be reduced during dosing to the largest possible value.

Note

The dosing rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Titrando, 814, 815, 855
Range

0.01 ... 166.00 mL/min