6 Configuration

tiamo 2.3 Patch 1

■■■■■■■■

1417

Name

Name of the colorimetric sensor.

Calibration date

Display of date and time of the last calibration entered after each auto-
matic calibration or manual entry.

Wavelength

Wavelength at which the calibration of the instrument has been carried
out.

Coefficient of determination

Display of the coefficient of determination r

2

, which is calculated on the

basis of the calibration function and the confidence interval and which is
between 0 und 1. 5 decimal places are displayed.

Confidence interval

Display of the confidence interval within which the measured values for
the determination of the calibration curve must lie.

c0

Calibration coefficient of zero order.

c1

Calibration coefficient first-order.

c2

Calibration coefficient second-order.

c3

Calibration coefficient third-order.

Additional columns from the properties of the colorimetric sensors can be
displayed with the menu item Edit

Column display....

Note

Lines that contain red entries will also show the line number with a red
background.

Table view
The table of the colorimetric sensors cannot be edited directly. With a click
on the column title the table can be sorted according to the selected col-