7 Instruments

tiamo 2.3 Patch 1

■■■■■■■■

1563

Light source always active

on | off (Default value: on)

If this check box is activated, then the LED on the photometer is always
on. Else the firmware takes care of switching it on before the measure-
ment or switching it off after the measurement.

Heater

Initial temperature

Initial temperature of the device (temperature of the cuvette). Entering an
initial temperature means that the heater will heat up to the set tempera-
ture when the device is switched on. off means that the heater will not be
switched on.

Range

20 ... 60 °C

Default value

30 °C

Selection

off

7.16.4

Properties - Adjustment

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

Adjustment

Transmission factor

Transmission factor, either determined via MEAS TMF or manually
entered here.

Range

0.0100 ... 20.0000

Default value

1.0000

Date

Shows time and date when the transmission factor was determined.

User

Shows the user logged in during the determination of the transmission
factor.

Method

Shows the method used to determine the transmission factor. If the trans-
mission factor was entered manually, manual is displayed.

Determination ID

ID of the determination during which the transmission factor was deter-
mined.