5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

636

■■■■■■■■

tiamo 2.3 Patch 1

Filling rate

Speed with which the dosing cylinder is to be refilled after the titration.
The maximum filling rate depends on the cylinder volume of the exchange
unit or dosing device used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

5.6.3.3.4.6

MET Ipol - Potentiometric evaluation

Tab: Method

MET Ipol Properties... Potentiometric evaluation

Command name

Name of the command.

Input

25 characters

Parameters for the potentiometric evaluation of the titration curve with
automatic equivalence point recognition are defined on this tab. One of
the following three evaluation methods can be chosen here:

Potentiometric evaluation

Selection

Evaluation without window | Evaluation with
measured value window (Upol)
| Evaluation
with volume window (mL)

Default value

Evaluation without window

Evaluation without window
With this option set, the two parameters EP criterion and EP recog-
nition
will be applied across the entire range of the titration curve.
Evaluation with measured value window (Upol)
With this option set, up to 9 regions (windows) can be defined on the
measured value axis. Only those equivalence points which lie within
these windows and additionally meet the parameters defined for each
window will be recognized. Per window only one EP will be recog-
nized. The defined windows with their parameters are shown in the
window table and can be edited with the following buttons: