5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

664

■■■■■■■■

tiamo 2.3 Patch 1

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (Upol)

EP criterion

Criterion for the recognition of equivalence points which is valid for all
windows. Equivalence points whose ERC is smaller than the set EP crite-
rion will not be recognized.

Titrino
Range

0.1 ... 99.9 µA

Default value

2.0 µA

[New]

Open the dialog window Measured value window # in which the
parameters for a new window can be entered (see Chapter 5.6.3.3.5.9,
page 670)
.

[Properties]

Open the dialog window Measured value window # in which the
parameters for the selected window can be edited (see Chapter
5.6.3.3.5.9, page 670)
.

[Delete]

Delete the window selected in the table.

Evaluation with volume window (mL)

[New]

Open the dialog window Volume window # in which the parameters
for a new window can be entered (see Chapter 5.6.3.3.5.10, page
671)
.