Metrohm tiamo 2.3 Patch 1 User Manual

Page 993

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

977

Selection

Device types | Titrando

Default value

Titrando

Sensor

Measuring input

Selection of the measuring input the sensor is connected to.

Titrando (without 851, 888, 890), 867
Selection

1 | 2

Default value

1

851, 855, 888, 890
Selection

1

Default value

1

852
Selection

2

Default value

2

Sensor

Selection of a sensor of type Metal electrode from the sensors config-
ured in the sensor table. The calibration data for the sensor will be adop-
ted for the determination.

Selection

Sensor name | Metal electrode

Default value

Metal electrode

U(pol)

The polarization voltage is the voltage applied to the polarized electrode
during an amperometric measurement.

Titrando, 855, 867
Range

-1250 ... 1250 mV (Increment: 25)

Default value

400 mV

Titrino
Range

-1270 ... 1270 mV (Increment: 10)

Default value

400 mV

Electrode check

on | off (Default value: off)

If this check box is enabled, then an electrode check will be carried out for
polarizable electrodes during the transition from an inactive normal status
to a measurement. A check is also made while doing so to ensure that the
electrode is properly connected and that no short-circuit is present.

Advertising