4 dos u - stop conditions, Stop conditions – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1205

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1189

5.6.6.3.5.4

DOS U - Stop conditions

Tab: Method

DOS U Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

The conditions for stopping the dosing are defined on this tab. If more
than one stop condition is in effect at the same time, then the criterion
which is fulfilled first will stop the dosing.

Stop volume

Stops when the given volume has been added after the start of the dos-
ing. The stop volume should be adapted to suit the measuring vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100.000 mL

Selection

off

Titrino
Range

0 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Range

0 ... 999999 s

Selection

off

Default value

off

off
Stops when the preset time has elapsed since the start of the dosing.

Filling rate

Rate at which the dosing cylinder is to be filled after dosing. The maxi-
mum filling rate depends on the cylinder volume of the exchange unit or
dosing device used. If the entered filling rate is too high for the selected
dosing device, it will automatically be reduced during dosing to the largest
possible value.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min