8 met ipol - additional measured values, Additional measured values – Metrohm tiamo 2.3 Patch 1 User Manual

Page 658

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

642

■■■■■■■■

tiamo 2.3 Patch 1

Gran evaluation

Note

This evaluation method is possible only with Titrando and 855.

Gran evaluation

on | off (Default value: off)

If this option is enabled, then titration curves will be evaluated in accord-
ance with the Gran Plot procedure(see Chapter 5.6.3.10.4, page 929).

Initial volume

Volume which must be present in the measuring vessel before the com-
mand is started.

Range

0.01 ... 9999.99 mL

Default value

50.00 mL

Lower limit

Value for the lower limit of the Gran evaluation range.

Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Upper limit

Value for the upper limit of the Gran evaluation range.

Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

5.6.3.3.4.8

MET Ipol - Additional measured values

Tab: Method

MET Ipol Properties... Additional measured values

Command name

Name of the command.

Input

25 characters

A maximum of 6 additional measured values can be defined on this tab
which can then be saved together in additional measured value columns
with the measured values present in the default settings:

Advertising