5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

818

■■■■■■■■

tiamo 2.3 Patch 1

5.6.3.5.3.5

KFT Upol - Titration parameters

Tab: Method

KFT Upol Properties... Titration parameters

Command name

Name of the command.

Input

25 characters

The parameters for the running of the titration are defined on this tab.

Titration direction

Selection of the titration direction.

Selection

+ | - | auto

Default value

auto

+
positive measured value modification
-
negative measured value modification
auto
The titration direction is determined automatically from the start meas-
ured value and the set endpoint.

Extraction time

The titration is not stopped until the extraction time has elapsed (even
when the EP has already been reached). The entry of an extraction time
may be advisable, e.g. for the titration of sparingly soluble samples.

Range

0 ... 999999 s

Temperature

Manually entered titration temperature. If a temperature sensor is connec-
ted and the Temperature measurement is set on the General/Hard-
ware
tab under Sensor to automatic or continuous, then the temper-
ature will be measured continuously. This value is used for temperature
correction in pH measurements.

Titrando, 855
Range

-20.0 ... 150.0 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C