6 editing rack properties (774), 7 properties - lift positions (774) – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1478

Advertising
background image

6.12 Subwindow Rack data

■■■■■■■■■■■■■■■■■■■■■■

1462

■■■■■■■■

tiamo 2.3 Patch 1

6.12.3.6

Editing rack properties (774)

Menu item: Configuration

Rack data Edit Properties…

Rack name

Shows the name of the rack attached. If no rack is in position then "-----"
will be shown.

Rack code

Shows the number of positions on the rack in position. The rack code cor-
responds to the arrangement of magnets on the base of the rack and is
read in by the Sample Processor in order to recognize the rack. If no rack
is in position then the display will be empty.

Number of positions

Shows the number of positions on the rack. If no rack is in position then
the display will be empty.

The parameters for the attached rack are defined on the following 2 tabs:

Lift positions
Defines the work, rinse, shift and special position for Tower 1.

Special beakers
Display of the special beaker for the rack in position.

6.12.3.7

Properties - Lift positions (774)

Dialog window: Configuration

Rack data Properties

Tower 1
Defines the lift positions for Tower 1. These apply for all rack positions
except those that are defined as (see Chapter 6.12.3.8, page 1463).

Work position

Work position for Lift 1. At this lift position the electrodes, stirrer and
buret tips are optimally positioned for work.

Range

0 ... 100 mm

Rinse position

Rinse position for Lift 1. This lift position is used for rinsing the electrodes.

Range

0 ... 100 mm

Shift position

Shift position for Lift 1. Each time that the rack shifts, the lift will move to
this position if it is located at a lower lift position. If the lift is located at a
higher lift position than that defined here, then the shifting will take place
at the current lift position. This means that the shift position must be

Advertising