5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

446

■■■■■■■■

tiamo 2.3 Patch 1

Conditioning

Automatic conditioning

on | off (Default value: on)

If this option is enabled, the conditioning will be started automatically at
the start of a determination for all commands with activated conditioning
(SET, KFT, KFC and BRC). After the track with the conditioning com-
mand has been finished, conditioning is started again automatically and
will be continued even after the end of the determination.

If this parameter is disabled, then conditioning will not be started until
the start of the command. Conditioning is not started again after the end
of the track with the conditioning command. This parameter must be
switched off for automation applications.

5.6.2.2.3

START - Application note

Tab: Method

START Properties... START 'Command name' Application

note

Command name

Name of the command.

Input

25 characters

The application note defined in the text window is displayed in the tab
Application note in the subwindow Live display of the program part
Workplace. The tab is opened automatically when the method is loaded.

The text editor with which the application note can be entered or

changed is opened with

or by double-clicking into the text field (see

Chapter 2.5.2, page 93).

5.6.2.2.4

START - Method variables (Table)

Tab: Method

START Properties... START 'Command name' Method varia-

ble

Command name

Name of the command.

Input

25 characters

The variables that are to be available within the method in other com-
mands (e.g. CALC command) under the designation MV.'Variable
name'
must be declared here. Either a sample data variable (entry in field
Assignment) or a direct value (entry in field Value) can be assigned to a
method variable.