5 Method

tiamo 2.3 Patch 1

■■■■■■■■

887

[Delete]

Delete the calculated measured value selected in the table.

Additional external measured values

Additional external measured values

on | off (Default value: off)

If this option is enabled, then a maximum of 3 new measured values can
be taken over and saved from existing measuring commands (i.e. MEAS)
which can also be displayed in curves under the designation Extern 1…3
and used in formulas as variables 'Command name.EX1…3'-

[New]

Open the dialog window External measured value # in which the
parameters for the external measured value can be entered (see Chapter
5.6.3.11.2, page 931)
.

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 931)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.8.6.10

STAT pH - Fixed endpoint evaluation

Dialog window: Method

STAT pH Properties... Additional evalua-

tions

Fixed endpoint evaluation [New]/[Properties] Fixed endpoint evalua-

tion #

Quantity

Selection of the fixed measured quantity to which the associated values
for the other quantities are to be interpolated from the list of measured
points.

Selection

Measured value | Time | Volume

Default value

Measured value

Fix value

Value of the fix endpoint.

Measured value
Range

-20.000 ... 20.000 pH

Time
Range

0.0 ... 999999.9 s