6 Configuration

tiamo 2.3 Patch 1

■■■■■■■■

1375

Statistics
Statistical data will only be shown if the titer has been assigned by the
titer method as a mean value of a multiple determination. If the titer has
been determined with a single determination or entered manually then no
statistics values will be shown.

Number of titer determinations (n)

Number of determinations used to determine the titer.

Abs. standard dev. (s abs)

Absolute standard deviation for the titer determination in the units of the
titer.

Rel. standard dev. (s rel)

Relative standard deviation for the titer determination in %.

Monitoring the titer validity
on | off (Default value: off)

If this option is activated, then the validity of the titer will be monitored.

Validity

Validity of the titer in days. If a value is entered here then the date for the
Next titer determination will be automatically adapted.

Range

0 ... 999 Days

Default value

999 Days

Next titer determination

Date on which the next titer determination is to be carried out. The date

can be selected by pressing

in the dialog window Select date (see

Chapter 2.5.1, page 92). After the date has been entered, the field Valid-
ity
will be adapted automatically.

Default value

   Date Titer det.+ 999 days

Message
The message options can only be edited if monitoring is switched on.

Message by E-mail

on | off (Default value: off)

The message is additionally sent to the e-mail address defined under [E-
mail]
if this option is activated. The message is sent in text format.